Zone 1 / Class I Div. 2
FOR TESTING
CURRENT LOOPS
TECHNICAL DATA
Ambient temperature -10 °C to +50 °C
Storage temperature -30 °C to +60 °C
Maximum voltage 28 Volt
Relative humidity
95 % (0 °C to +30 °C);
75 % (+30 °C to +40 °C);
45 % (-40 °C to +50 °C)
Power supply 1x 6LR61, type approved
Operating time 18 hours typical, at 12 mA
Dimensions (HxWxD) 164 x 75 x 47 mm (with holster)
Weight 350 g (with holster)
SUMMARY SPECIFICATIONS (18 °C TO 28 °C. ONE YEAR)
Function Range Resolution Accuracy
Voltage measure 0 to 28 V 0.001 V ±(0.015 % v. MW. +2 Digits)
mA measure 0 to 24 mA 0.001 mA ±(0.015 % v. MW. +2 Digits)
mA source
1
0 to 24 mA 0.001 mA ±(0.015 % v. MW. +2 Digits)
mA simulate
2
0 to 24 mA 0.001 mA ±(0.015 % v. MW. +2 Digits)
Loop supply 24 V n. z. 24 V ± 1 V DC
Temperature Coefcient, -10 to 18 °C, 28 to 55 °C: ± 0.005 % of range per °C;
1
Max load, 700 Ohms at 20 mA
2
Max applied voltage for simulation, 28 V